Cocoon VSD Occluder

Cocoon VSD Occluder

Platinum nanocoating technology

Nitinol wire used in the Cocoon VSD Occluder is coated with Platinum atoms using nano fusion technology by plasma deposition

Raw nitinol

Electropolished

Platinum nanocoating

Platinum nanocoating makes the Cocoon Septal Occluder

Key highlights

*Recapture and redeployment is possible only if delivery cable is securely connected to the disc.

Design

  • Left ventricular disc diameter (Ø1)
  • Right ventricular disc diameter (Ø2)
  • Waist diameter (Ø3)
  • Waist length (L)

Thickness of the septum is considered for selecting the length of the device

Technical Specifications

Product CodeLeft ventricular disc diameter (Ø 1)
(mm)
Right ventricular disc diameter (Ø 2 )
(mm)
Waist diameter (Ø3)
(mm)
Waist length (L)
(mm)
Recommended for VSD sizeAccessory catalog number
COV040410103.2543 mmCOV6F or COVII6F
COV060412125.2543.1 to 5 mmCOV6F or COVII6F
COV080414147.2545.1 to 7 mmCOV6F or COVII6F
COV100416169.2547.1 to 9 mmCOV7F or COVII7F
COV1204181811.2549.1 to 11 mmCOV7F or COVII7F
COV04071010472 mmCOV6F or COVII6F
COV06071212672.1 to 4 mmCOV6F or COVII6F
COV08071414874.1 to 6 mmCOV6F or COVII6F
COV100716161076.1 to 8 mmCOV7F or COVII7F
COV120718181278.1 to 10 mmCOV7F or COVII7F
COV041010104102 mmCOV6F or COVII6F
COV061012126102.1 to 4 mmCOV6F or COVII6F
COV081014148104.1 to 6 mmCOV6F or COVII6F
COV1010161610106.1 to 8 mmCOV7F or COVII7F
COV1210181812108.1 to 10 mmCOV7F or COVII7F

Usable lengths of compatible accessories (cm)

Usable Length (cm)

6F7F
Sheath8383
Dilator8989
Loader99

Delivery Cable

Catalog numberDescriptionDiameter (inch)Length (cm)
Delivery cable 0.030 140
CDCVDelivery cable sheath 125

Preclinical Studies

Ventricular Septal Defect was created in swine models and subsequently implanted with a Cocoon VSD Occluder

Explants at 30 days show the device incorporated with tissues

Preclinical studies

Echo images of the swine at a day 1 and day 30 showing the complete closure of the defect

Caution

This product is intended for use by or under the direction of a physician. Prior to use, refer to the “Instructions for use” supplied with these devices for indications, contraindications, side effects, suggested procedure warnings and precautions. As part of our continuous product development policy, we reserve the right to change product specifications without prior notification. Information contained herein is for distribution outside the USA & Japan.